Intended Use
These sensors are designed to be used to monitor the partial pressure of oxygen in anaesthesia (not including xenon), critical care, incubators and general oxygen monitors
An 'Instruction For Use' leaflet (RM945 Issue 1.0) is included with each sensor.
Stablisation Time
Allow at least 15 minutes to stabilise in the instrument before calibration or refer to manufacturer's instructions.
Cleaning and Sterilisation
In case of contamination the sensor may be cleaned with disculled water and allowed to dry naturally. The sensor is not suitable for sterilisation by steam or exposure to chemicals such as ethylene oxide or hydrogen peroxide.
Calibration Interval
These sensors are designed to have minimal drift over their useful lifetime. For maximum accuracy however, they should be calibrated before each use.
If the Sensor is Dropped
If a sensor is dropped, then it should be placed in quarantine for 24 hours and a follow-up check made by a 2 point calibration.
Mechanical Installation
When installing the sensor, it must only be screwed in hand-tight and a gas tight seal ensured. Spanners and similar mechanical aids may not be used, as excessive force may damage the sensor thread.